LCMS data
| Crude drug latin name | CORNI FRUCTUS | ||||||||||
| Sample | NIB-0480 | ||||||||||
| CHROM_NAME | ESI-Orbitrap(-) | ||||||||||
| CHROM_TYPE | LC | ||||||||||
| LC MACHINE TYPE | ThermoFisher Scientific Ultimate 3000 | ||||||||||
| MS MACHINE TYPE | ThermoFisher Scientific LTQ Orbitrap Elite | ||||||||||
| Column | FILTER | ODS | |||||||||
| LENGTH | 100mm | ||||||||||
| DIAMETER | 2.1mm | ||||||||||
| PARTICLES_SIZE | 2.6 μm | ||||||||||
| MANUFACTURER | ChromaNik Technologies | ||||||||||
| PRODUCT_NAME | SunShell C18 | ||||||||||
| LC Measurement condition | MOBILE_PHASE | A = 10mM ammonium formate in ultrapure water; B = 0.01% formic acid in acetonitrile | |||||||||
| LC_G_PROGRAM | B = 1% - 1 % (1 min) - 40 % (12 min) - 100 % (14 min) - 100 % (16 min) | ||||||||||
| COLUMN_TEMP | 40℃ | ||||||||||
| FLOW_RATE | 0.3 mL/min | ||||||||||
| MS Measurement condition | Ionization mode | negative | |||||||||
| Desolvationtemperature | 275℃ | ||||||||||
| Needle voltage | 3000V | ||||||||||
| Velocity of Spectrum | 10 spec./sec. | ||||||||||
| Range of m/z | 100-2000 | ||||||||||
| Collision gas | He | ||||||||||
| TIC_JCAMP_FILE | NIB-0480(-).jdx | ||||||||||
| 3D(PDA) data | |||||||||||
| UV data | |||||||||||
| |||||||||||
| MEMO | sample:SANSHUYU (Cornus Fruit) water ext. 1mg/mL (water/methanol 1:1) Injection Volume: 5µL サンシュユ 水抽出エキス 1mg/mL (水/メタノール 1:1溶液) 注入量: 5μL | ||||||||||
| MS data list | |||||||||||
| |||||||||||