| Crude drug latin name
| MAGNOLIAE FLOS |
| Sample
| NIB-0918 |
| CHROM_NAME
| ESI-Orbitrap(-) |
| CHROM_TYPE
| LC |
| LC MACHINE TYPE
| Shimadzu 30A |
| MS MACHINE TYPE
| ThermoFisher Scientific Orbitrap Velos Pro |
| Column
| FILTER
| ODS |
| LENGTH
| 150mm |
| DIAMETER
| 2.0 mm |
| PARTICLES_SIZE
| 3 µm |
| MANUFACTURER
| YMC |
| PRODUCT_NAME
| YMC-Triart C18 |
| LC Measurement condition
| MOBILE_PHASE
| A = 0.1% formic acid in ultrapure water; B = acetonitrile |
| LC_G_PROGRAM
| B = 5% - 5 % (2 min) - 95 % (18 min) |
| COLUMN_TEMP
| 40℃ |
| FLOW_RATE
| 0.3 mL/min |
| MS Measurement condition
| Ionization mode
| negative |
| Desolvationtemperature
| 450 ℃ |
| Needle voltage
| 2500V |
| Velocity of Spectrum
| |
| Range of m/z
| 100-2000 |
| Collision gas
| N2 |
| TIC_JCAMP_FILE
|  NIB-0918(-).jdx |
| 3D(PDA) data
|  |
| UV data
|
| UV chromatogram
|  |
| Wavelength(nm)
| 190 - 400 |
| Memo
| measured on a PDA |
|
| MEMO
| sample: SHIN-I (Magnolia Flower) water ext.
Injection Volume: 10 μL |
| MS data list
|
|
|